Interested in RNA chemical biology and structure, even better if some NMR is involved
I also tested the 15N sofast-HMQC library pp sfhmqcf3gpph, with 128 increments, 125 ms D1, 71 ms AQ, and that completed normally. So only the 13C pp sf_mehmqcgpph seems to be affected
I also tested the 15N sofast-HMQC library pp sfhmqcf3gpph, with 128 increments, 125 ms D1, 71 ms AQ, and that completed normally. So only the 13C pp sf_mehmqcgpph seems to be affected
The experiment was already done with digital acquisition mode. Using baseopt also lead to an early abort + the IPSO error below that forced a restart.
The experiment was already done with digital acquisition mode. Using baseopt also lead to an early abort + the IPSO error below that forced a restart.
- TS 3.8.0
- D1 of 200 ms, AQ 50 ms in the direct dimension
- indirect TD of 160, typically aborts after 50-70
- With a lower NS, e.g. 8 instead of 720, the experiment goes to completion
- A combination of low NS and shorter D1 (e.g. 8 scans with 100 ms) also errors out again
- TS 3.8.0
- D1 of 200 ms, AQ 50 ms in the direct dimension
- indirect TD of 160, typically aborts after 50-70
- With a lower NS, e.g. 8 instead of 720, the experiment goes to completion
- A combination of low NS and shorter D1 (e.g. 8 scans with 100 ms) also errors out again